Using COD to Measure Lost Product
Prepared by:
Lynn G. Turner and Roy E. Carawan
Extension Food Science Specialists
North Carolina State University
Published by: North Carolina Cooperative Extension
Service
Publication Number: CD-38
Last Electronic Revision: March 1996 (JWM)
Good processing plants produce various types of waste materials (solid, liquid, and gaseous) during routine operations. Indiscriminate or improper disposal of these waste materials or discharge of improperly treated waste materials can harm the environment and can result in surcharges, fines, or civil and criminal penalties.
As environmental regulations become stricter and the food industry expands, it becomes important for processors to have methods for sampling waste streams. Routine sampling and testing will enable a food processor to determine which activities and products produce the greatest waste streams.
What Are BOD5 AND COD?
Chemical oxygen demand (COD) and biochemical oxygen demand (BOD5) are measurements commonly used to determine water quality. Many of the wastes in the effluent from food processing plants are organic compounds, primarily lost product. As these substances oxidize or stabilize, they combine with some of the oxygen dissolved in the water. The amount of oxygen used is therefore a good indicator of the amount of organic waste present. The BOD5 and COD values for three dairy products are shown in the table above. The values indicate the amount of oxygen (in milligrams per liter of product) needed to oxidize or stabilize these products when they appear in wastewater.
COD and BOD5 are of special interest to the food processing industry as they can be a useful tool for indicating lost product and wasteful practices. High BOD5 and COD levels indicate increased amounts of product lost to the waste stream.
Greater numbers of food processing plants are taking steps to reduce waste and save water because of rising treatment costs and increased environmental standards. BOD5 and COD measurements can help isolate areas of severe waste.
Why COD?
Although regulatory agencies require the monitoring and reporting of BOD5 levels, the COD test has several advantages for the plant's operation staff. Time (speed) is the major advantage: a COD 2-hour reflux test is standard. A COD test of easily oxidized waste may take even less time. The BOD5 test is time consuming, requiring a standard 5-day incubation period. The rapid test results of the COD procedure provide an advantage to the processor monitoring daily waste production and wastewater discharge.
Another advantage of the COD test is that strong oxidizing conditions are somewhat independent of variations in experimental conditions and procedures.This is not true for the BOD5 test, which is sensitive to test conditions and can vary depending upon the seed (microbial) culture used. In addition, the BOD5 test, unlike the COD test, does not measure biologically resistant compounds.
Values of BOD5 and COD for Food Plant Wastewaters
Type of Processor BOD5 COD BOD5/COD
(mg/l) (mg/l)
Poultry processor 1,306 1,581 0.83
Frozen bakery products 3,200 7,000 0.46
Meat specialties 530 900 0.59
Jams and jellies 2,400 4,000 0.60
Meat packing 1,433 2,746 0.52
Dairy processing 2,700 4,700 0.57
Gulf shrimp canning 1,081 2,296 0.47
Are COD and BOD5 Interrelated?
At any point in a particular food processing operation, the relationship between BOD5 and COD is fairly consistant. The ratio's of these two measures varies widely with the type of product, however. Typical BOD5 and COD values for the wastewater from seven types of food processing operations are shown in the table on this page. FOr example, a bakery product having a COD of 7,000 mg/l would have a corresponding BOD5 of 3,200 mg/l. The ratio of BOD5 to COD , shown in the right column, ranges from .46 to .80. The ratio also depends on where the measurements are made in the processing operation or in the wastewater treatment process.
COD values are greater than BOD5 values beacuase of the nature of the measurement procedure. With the dichromate refluxing procedure used to measure COD, almost all organic compounds will oxidize. With the BOD5 measurement procedure, some of these compounds do not fully oxidize, making the oxygen demand lower. The BOD5 value may be much lower than the COD value when a substantial amount of bilogically resistant organic matter is present. In addition, a few chemical interferences-primarily from chlorides and certain nitrogen compounds-can affect the test results. For furtherinformation, see "Elimination of Interference."
Dichromate Reflux Technique Standard Method.
Equipement Required
- 500-milliliter (ml) Erlenmeyer flask with standard (24/40) tapered glass joints
- Friedrichs reflux condensers (12-inch) with standard (24/40) tapered glass joints
- Electric hot plate or six-unit heating shelf
- Volumetric pipets (10, 25, and 50-ml capacity)
- Buret, 50 ml - 0.1 ml accuracy
- Buret stand and clamp
- Analytical balance, accuracy 0.001 gram (g)
- Spatula
- Volumetric flasks (1,000 ml capacity)
- Boiling beads, glass
- Magnetic stirrer and stirring bars
Chemicals Required
- Potassium dichromate (K2Cr2O7) 0.25 N
- Sulfuric acid (H2SO4) silver sulfate (Ag2SO4) solution
- Mercuric sulfate (HgSO4) crystals
- Ferrous ammonium sulfate (FAS) [Fe(NH4)2(SO4)2], approximately 0.01 N
- Ferroin indicator (1, 10-phenanthroline and ferrous ammonium sulfate)
Caution: In carrying out the following procedures, use proper
safety measures, including protective clothing, eye protection,
and a fume hood. Reagents containing heavy metals (HgSO4 and Ag2SO4)
should be disposed of as toxic wastes.
Chemical Preparation
- Dissolve 12.259 g of oven-dried (primary standard grade dried 103oC
to a constant weight) potassium dichromate in distilled water and dilute to
1-liter volume in a volumetric flask.
- Add 22 g of reagent grade silver sulfate to a 4-kg bottle of
concentrated sulfuric acid (H2SO4) and mix until the silver sulfate goes
into solution.
- Use 1 g of mercuric sulfate (HgSO4) to complex 100 mg chloride
(2,000 mg/l).
- Dissolve 1.485 g of 1,10-phenanthroline monohydrate and 0.695 g
of ferrous ammonium sulfate heptahydrate in distilled water and dilute
to approximately 100 ml. (Alternatively, this indicator may be purchased
from most scientific supply houses.)
- Dissolve 39 g reagent grade ferrous ammonium sulfate hexahydrate
in distilled water. Add 20 ml of concentrated sulfuric acid (H2SO4).
Cool and dilute to exactly 1 liter in a volumetric flask using distilled
water. The ferrous ammonium sulfate (FAS) titrant must be standardized
daily by the following procedure. Dilute 10 ml of standard potassium
dichromate (K2Cr2O7) solution to 100 ml with distilled water. Slowly
add 30 ml of concentrated sulfuric acid and cool to room temperature.
Titrate with ferrous ammonium sulfate titrant, using 2 to 3 drops
(0.10 to 0.15 ml) of ferroin indicator.
Normality of FAS = (ml K2Cr2O7)(0.25)
ml FAS required
The deterioration of FAS can be decreased if it is stored in a
dark bottle.
Procedure
- Place a 50-ml sample or an aliquot diluted to 50 ml
in a 500-ml refluxing flask. The blank is prepared using 50
ml of distilled water. This is a precise measurement and a 50-ml
volumetric pipet should be used. (Refer to "COD Range and Sample
Size" below for dilution)
- Add 5 to 7 glass boiling beads.
- Add 1 g of mercuric sulfate (HgSO4), 5 ml of
concentrated sulfur-ic acid, mercuric sulfate solution, and
mix until the HgSO4 is in solution. The function of the mercuric
sulfate is to bind or complex chlorides. One gram may not be
required if the chloride concentration is low.
(Refer to "Elimination of Interference." p.4
Caution: Always add acid slowly down the side of the flask
while mixing to avoid overheating. It may be necessary to use
gloves because of the heat generated.
- Add 25 ml of 0.25 N potassium dichromate (K2Cr2O7)
and mix.
- Add while mixing an additional 70 ml of concentrated
sulfuric acid-silver sulfate solution.
- After thorough mixing, attach the flask to the reflux
condenser, apply heat, and reflux for 2 hours. Refluxing time
can be decreased depending on the ease of oxidation of organic
materials. This time may be determined by refluxing for periods
from 15 minutes to 2 hours and comparing the results.
- A reagent blank containing 50 ml of distilled water
treated with the same reagent as the sample should be refluxed
with each set of samples.
- Cool the apparatus to room temperature after the
refluxing period. Wash down the interior of the condenser
and flask twice with approximately 25-ml portions of distilled
water.
- Remove flask from the condenser and dilute to a final
volume of approximately 350 ml with distilled water.
- Add 4 to 5 drops of ferroin indicator and a magnetic
stirring bar.
- Place flask on a magnetic stirrer and rapidly titrate
with 0.1 N ferrous ammonium sulfate to the first red brown
endpoint.
Caution: Use care in titration. The endpoint is very
sharp and may be reached rapidly.
- Formula to determine COD
COD (mg/l) = (a-b)(N) x 8,000
sample size (ml)
a = ml Fe(NH4)2(SO4)2 used for blank
b = ml Fe(NH4)2(SO4)2 used for sample
N = normality of ferrous ammonium sulfate (FAS) Fe(NH4)2(SO4)2
ml sample = the actual volume of sample used before dilution
Sources of Error
- The largest error is caused by using a nonhomogeneous sample. Every effort should be made to blend and mix the sample so that solids are never excluded from any aliquot.
- Always use the largest sample practical and use the largest glassware that is in keeping with good laboratory practice.
- Use volumetric flasks and volumetric pipets with a large bore.
- The K2Cr2O7 oxidizing agent must be precisely measured. Use a volumetric pipet and use the same one each time if possible.
- When titrating, be certain that the buret is clean and free of air bubbles.
- Always read the bottom of the meniscus and position the meniscus at eye level.
COD Range and Sample Size
COD Range(mg/l) 50-800 100-1500 240-3700 480-7500 1200-18800 2400-3700 2400-375000
Volume of
Sample (ml) 50 25 10 5 2 1 0.1
All samples high in solids should be blended for 2 minutes at high speed and stirred when an aliquot is taken. Sample volumes less than 25 ml should not be pipetted directly, but serially diluted and then a portion of the diluent removed:
- 500 ml of sample diluted to 1,000 ml = 0.5 ml sample/ml of diluent, .: 50 ml of diluent = 25 ml of sample.
- 100 ml of sample diluted to 1,000 ml = 0.1 ml sample/ml diluent, .: 50 ml of diluent = 5 ml of sample.
Elimination of Interference
One gram of mercuric sulfate (HgSO4) will complex 100 mg of chloride in a 50-ml sample (2,000 mg/l). For samples higher in chloride more HgSO4 should be used in the ratio of 10:1 HgSO4.
Interference from nitrites can be prevented by the addition of 10:1 ratio of sulfamic acid:nitrite. The addition of the silver sulfate (AgSO4) concentrated sulfuric acid (H2SO4) refluxing acid will aid in the oxidation of some organic nitrogen compounds, but aromatic hydrocarbons and pyridine are not oxidized to any appreciable amount.
HACH Semimicro-Alternate Procedures
Equipment and Chemicals Required
HACH COD test system kit including:
Procedure HACH Semimicro
WARNING
- Vials contain hazardous materials (sulfuric acid and mercuric salts) and are heated to 150oC (302oF). The aluminum block heater could cause severe burns if touched, and a glass vial could burst. Wear protective clothing (gloves, goggles, or face mask) when performing the test. Wipe up spills immediately and rinse any skin area contacted with cool water.
- Bare terminals on the reactor could cause severe shock if touched.
Always disconnect power before removing the reactor cover.
- Place approximately 500 ml of sample in a clean blender bowl
and homogenize at high speed for 2 minutes. This ensures a uniform
distribution of suspended solids and improves both accuracy and repeatability.
- Preheat COD reactor to 150oC. Place a towel around a COD digestion
reagent vial of desired range and remove the cap. Carefully pipet 2.00 ml
of sample with a large-bore pipet into the vial while holding the vial at
a 45-degree angle directed away from you.
Caution: Do not use vials that have reagent spilled or leaked because
the test accuracy will be affected and reagents may be hazardous to the
skin.
- Using a dry towel wrapped around the vial, cautiously replace
and tighten cap. Hold vial by the cap over an empty sink and invert
several times to mix contents.
Caution: Vial will become hot during mixing procedure and the towel
will decrease heat transfer.
- Have plastic shield in place on heater block and insert vial in
preheated COD reactor.
- Prepare a reagent blank by substituting 2.00 ml of distilled or
demineralized water for the wastewater sample and follow steps 1 through 4.
- Heat vials at 150oC for 2 hours. Turn off the reactor and allow
vials to cool to 120oC or less (requires approximately 20 minutes).
Invert each vial several times while still warm.
- Place vials in cooling rack and allow to cool to touch.
- Measure COD by the titrametric or colorimetric method. One
blank from same vial lot should be tested with each set of samples.
- COD standard solutions can be used to replace wastewater samples
to monitor accuracy.
Colorimetric Measurement (0 to150 mg/l COD)
- Place the COD vial adapter in the cell holder of the spectrophotometer
with the raised marker to the left. Insert the vial containing the reagent
blank into the adapter and install the cover. Insert the oxygen demand,
chemical low-range meter scale into the meter and adjust the wavelength
dial to 600 nm.
- Set mode switch to LEFT SET and adjust the LEFT SET control to align
the meter needle with the extreme left mark of the meter scale arc.
- Set the mode switch to NORM and adjust the RIGHT SET control for a
meter reading of 0 mg/l.
- Place the vial containing the test sample in the adapter, replace
light shield, and read mg/l from the meter scale.
Modification of Alternative Procedures
Modification of alternative procedures can result in a cost savings. Changes that produce similar results but utilize available equipment can show a sizable savings in the initial or start-up costs. The following modifications can be utilized for cost or time savings once a procedure is found that produces desired results. Modifications include:
- Using an oven that maintains 150oC ±2oC instead of purchasing a manufacturer's heating control.
Caution: Maintain precise temperature and ensure a pressure tight seal on the digestion vials. (A shield should be provided to protect against explosion when heating vials).
- Snap-off glass vials can be heat sealed with a propane torch.
Caution: Prevent boiling the digestion mixture by applying heat to tip or end of sealing neck.
A colorimetric procedure may be accomplished with the use of any spectrophotometer that can produce a 600-nm wavelength and has a 1-cm light path. A standard curve can be prepared by digestion of known COD value samples and plotting concentration versus percent transmittance. A sample of 850 mg of potassium acid phthalate dried to a constant weight at 10oC diluted to 1 liter with carbon dioxide (CO2) free distilled water will produce 1,000 mg/l COD.
A titrametric procedure can be accomplished by quantitatively transferring the contents of the digestion vials to an Erlenmeyer flask and diluting it to approximately 200-ml final volume with distilled water. Titrate to the “ferroin” endpoint with 0.015 N ferrous ammonium sulfate.
References
American Public Health Association. 1985. Standard Methods for the Examination of Water Waste, 16th ed. American Public Health Association, New York.
Hach Company. 1987. Analytical Procedures, Oxygen Demand, Chemical. Loveland, Colorado.
Sawyer, C. N. 1960. Chemistry for Sanitary Engineers. McGraw-Hill, New York.
Sawyer, C. N., and P. L. McCarty. 1978. Chemistry for Environmental Engineering, 3rd ed. McGraw-Hill, New York.
The use of brand names in this publication does not imply
endorsement of the products or services named or criticism
of similar ones not mentioned.
Distributed in furtherance of the Acts of Congress of May 8 and June 30,
1914. Employment and program opportunities are offered to all people
regardless of race, color, national origin, sex, age, or disability. North
Carolina State University, North Carolina A&T State University, U.S.
Department of Agriculture, and local governments cooperating.
CD-38